Introduction to KONICA MINOLTA CM-36dGV Vertical Spectrophotometer
The Konica Minolta CM-36dGV vertical color and gloss spectrophotometer is a stationary benchtop upright spectrophotometer. It integrates a d/8 optical geometry along with a gloss sensor, enabling simultaneous measurement of color and gloss while maintaining backward data compatibility with the long-time best-selling KONICA MINOLTA CM-3600A series (when measuring in SCI/LAV mode).
Its superior measurement accuracy supports digitized quality control across global supply chains. The KONICA MINOLTA CM-36dGV spectrophotometer series is designed with a vertical, bottom-port configuration and an integrated gloss measurement module.
Key Features of Konica Minolta CM-36dGV
-
Simultaneous color and gloss measurement with high Inter-Instrument Agreement (IIA).
-
Colorimetric: $\Delta E^*ab < 0.12$ (LAV/SCI; average of 12 BCRA Series II color tiles).
-
Gloss: $\pm0.2$ GU (MAV; 0 – 10 GU).
-
Wavelength Analysis & Adjustment (WAA) function.
-
Excellent Inter-Model Agreement (IMA) with the legacy CM-3600A series.
-
Status panel displaying all device settings and operational conditions.
-
4 different measurement areas (apertures):
-
LAV ø25.4 mm
-
LMAV ø16 mm
-
MAV ø8 mm
-
SAV ø4 mm
-
-
Sample preview viewfinder function.
-
Vertical design optimizes handling for flexible sheet-like samples such as textiles, fabrics, and paper.
Technical Specifications of KONICA MINOLTA CM-36dGV
- Color
- Illumination/Viewing system:
- Reflectance:
- di:8°, de:8° (diffuse illumination, 8° viewing angle), switchable SCI (Specular Component Included) / SCE (Specular Component Excluded)
- Complies with CIE No. 15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition c standards
- Transmittance:
- di:0°, de:0° (diffuse illumination, 0° viewing angle)
- Complies with CIE No.15 condition g, ASTM E1164, DIN 5033 Teil 7, and JIS Z 8722 standards
- Reflectance:
- Integrating sphere size: Ø152 mm (6 inches)
- Detector: Dual 40-element silicon photodiode array
- Spectral separation device: Diffraction grating
- Wavelength range: 360 nm to 740 nm
- Wavelength pitch: 10 nm
- Half bandwidth: Approx. 10 nm
- Reflectance range:
- 0 to 200%
- Output resolution: 0.01%
- Light source: 3 pulsed Xenon lamps (2 lamps equipped with UV-cut filters)
- Measurement/Illumination area:
- LAV: Ø25.4 mm / Ø30.0 mm
- LMAV: Ø16.0 mm / Ø20.0 mm
- MAV: Ø8.0 mm / Ø11.0 mm
- SAV: Ø4.0 mm / Ø7.0 mm
- Transmittance: Ø17.0 mm / Ø24.0 mm
- Repeatability:
- Colorimetric value: Standard deviation within $\Delta E^*ab \le 0.02$
- Spectral reflectance: Standard deviation within 0.1%
- (When a white calibration plate is measured 30 times at 10-second intervals after white calibration)
- Inter-instrument agreement: $\Delta E^*ab \le 0.12$ (Based on the average of 12 BCRA Series II color tiles; LAV/SCI; compared to values measured with the master body under Konica Minolta standard conditions)
- UV setting:
- 100% / 0% / Adjusted
- Instantaneous Numerical UV Adjustment without mechanical filter movement; includes 400 nm and 420 nm UV cutoff filters
- Illumination/Viewing system:
- Gloss
- Measurement angle: 60°
- Light source: White LED
- Detector: Silicon photodiode
- Measurement range:
- 0 to 200 GU
- Resolution: 0.01 GU
- Measurement area:
- MAV (for LAV/LMAV/MAV color apertures): Ellipse Ø10.0 x 8.0 mm
- SAV (for SAV color aperture): Ø3.0 mm
- Repeatability:
- 0 to 10 GU: within 0.1 GU
- 10 to 100 GU: within 0.2 GU
- 100 to 200 GU: within 0.2%
- (When measured 30 times at 10-second intervals)
- Inter-instrument agreement:
- 0 to 10 GU: ± 0.2 GU
- 10 to 100 GU: ± 0.5 GU
- (MAV. Compared to values measured with the master body under Konica Minolta standard conditions)
- Geometry: JIS Z 8741 (MAV), JIS K 5600, ISO 2813, ISO 7668 (MAV), ASTM D523-08, ASTM D2457-13, and DIN 67530
- Measurement time of CM-36dGV:
- Approx. 3.5 seconds (SCI + SCE measurement)
- Approx. 4.0 seconds (SCI + SCE + Gloss measurement)
- Approx. 2.5 seconds (Transmittance)
- Approx. 3.0 seconds (UV-cut/UV-adjusted; SCI or SCE)
- Minimum measurement interval:
- Approx. 4.0 seconds (SCI + SCE measurement)
- Approx. 4.5 seconds (SCI + SCE + Gloss measurement)
- Approx. 3.0 seconds (Transmittance)
- Approx. 4.0 seconds (UV-cut/UV-adjusted; SCI or SCE)
- Sample preview function:
- Utilizes internal camera
- (Images can be viewed/captured using optional color software, such as SpectraMagic NX Version 3.2 or later)
- Internal performance check: WAA (Wavelength Analysis & Adjustment) technology
- Interface: USB 2.0
- Automatic target mask detection: Yes
- Power supply: Dedicated AC adapter
- Operating temperature/humidity range: 13 to 33°C, relative humidity: 80% or less (at 35°C) non-condensing
- Storage temperature/humidity range: 0 to 40°C, relative humidity: 80% or less (at 35°C) non-condensing
- Dimensions (W x H x D): Approx. 300 x 677 x 315 mm
- Weight of CM-36dGV: Approx. 14.0 kg












There are no reviews yet.