Konica Minolta CM-3700A Spectrophotometer
The Konica Minolta CM-3700A is a flagship benchtop spectrophotometer, renowned for its ability to deliver exceptionally accurate and reliable measurement results. Engineered with flexible aperture selections, an advanced optical design, and a large-diameter integrating sphere, this instrument easily satisfies and exceeds stringent international standards. The implementation of a high-performance pulsed Xenon flash during measurement ensures optimal precision.
Users are granted the operational flexibility to switch the measurement area using standard aperture selections of 25.4mm, 8mm, and 3x5mm, catering to the specific needs of each application. The Konica Minolta CM-3700A also comes equipped with a sample holder featuring an extended arm, enabling precise measurements at the center of specimens up to A4 dimensions, while a specialized damper mechanism prevents sample damage caused by accidental forceful closures.
Key Features of the KONICA MINOLTA CM-3700A
- The CM-3700A is equipped with state-of-the-art technology to guarantee maximum precision and long-term traceability.
- Full Windows compatibility allows for effortless system control via PC connection using corresponding color management software.
- Provides high measurement flexibility accommodating versatile aperture area dimensions.
- Enables seamless evaluation of reflective, translucent, and fluorescent color specimens thanks to automated SCI/SCE switching, diffuse illumination geometry, and integrated UV cutoff filters.
- Complies strictly with numerous international standard specifications including CIE, ISO, ASTM, DIN, and JIS.
- The CM-3700A utilizes a high-intensity Xenon flash source to acquire highly precise data even on deep, dark, or highly saturated color samples.
- Standard USB interface ensures rapid and stable data connectivity with host PCs.
- The integrated sample holder features a smooth dampening mechanism to protect specimens from surface marking during clip closure.
- The physical power switch is shielded by a protective frame assembly to prevent accidental system power-offs during routine operations.
Technical Specifications of CM-3700A
Illumination / Viewing System:
- Reflectance: Switchable di:8° and de:8° (diffuse illumination, 8° viewing angle).
- Reflectance modes support both SCI (Specular Component Included) and SCE (Specular Component Excluded) simultaneous calculation.
- The CM-3700A complies fully with CIE No. 15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition c standard specifications.
- Transmittance: di:0° and de:0° (diffuse illumination, 0° viewing angle).
- Complies fully with condition g of CIE No. 15, ASTM E1164, DIN 5033 Teil 7, and JIS Z 8722 standard specifications.
Photodetector:
- Dual silicon photodiode array sensor with a planar structure.
Spectral Parameters:
- Wavelength range spans from 360 nm to 740 nm.
- Standard 10 nm spectral measurement pitch.
- Optical half bandwidth averages approximately 14 nm.
- Photometric range: 0 to 200%; readout resolution: 0.001%.
Illumination Source & Cycle Interactivity:
- Two pulsed Xenon arc lamps.
- The CM-3700A features a minimum measurement interval cycle time of 3.0 seconds.
Illumination / Measurement Aperture Areas:
- Reflectance is switchable between LAV, MAV, and SAV configurations:
- SAV: 5 × 7 mm illumination area / 3 × 5 mm measurement aperture area.
- MAV: Ø11 mm illumination area / Ø8 mm measurement aperture area.
- LAV: Ø28 mm illumination area / Ø25.4 mm measurement aperture area.
- Transmittance: Illumination spot size approx. Ø20 mm.
Measurement Repeatability & Instrumental Traceability:
- When a standard white calibration plate is measured 30 times at 10-second intervals after white calibration, spectral reflectance exhibits a standard deviation within 0.05% and a chromaticity tolerance within $\Delta E^*ab \le 0.005$.
- When evaluating a standard black calibration tile featuring 1% reflectance, the spectral reflectance standard deviation ranges within 0.02% (from 380 to 740 nm) and within 0.04% (at 360 and 370 nm), with a chromaticity tolerance within $\Delta E^*ab \le 0.05$.
- Inter-Instrument Agreement (IIA) for the LAV aperture operates within an average of $\Delta E^*ab \le 0.08$ and a maximum limit of $\Delta E^*ab \le 0.3$.
Numerical UV Calibration & Specimen Chamber:
- UV parameter adjustment is controlled via host computer software interface; features an integrated 400 nm UV cutoff filter.
- Accommodates a maximum sample thickness of 50 mm, with unlimited sample length when the chamber cover is left open.
- The sample transmission unit transmission chamber is fully detachable.
Connectivity Ports & Power Requirements:
- Standard USB 1.1 communication interface port.
- Universal power supply accommodating AC 100V to 240V, 50/60Hz lines.
Environmental Operating & Storage Conditions:
- Operating temperature range: 13 to 33°C.
- Operating relative humidity limit: 80% or less non-condensing (at 33°C).
- Storage relative humidity limit: 80% or less non-condensing (at 35°C).
- Storage temperature range: 0 to 40°C.
Physical Dimensions & Weight:
- Outer physical dimensions of the instrument: 271 × 274 × 500 mm.
- Net weight of the benchmark assembly: 18 kg.
Standard Included Accessories:
- White Calibration Plate.
- Target Masks for three aperture size profiles: 3 × 5 mm, Ø8 mm, and Ø25.4 mm.
- Zero Calibration Box.
- Dedicated AC Power Adapter assembly.
- 3-meter USB Connection Cable.
- Accessory Storage Case.
- Heavy-duty Fabric Dust Cover.
Optional Accessories for the CM-3700A:
- SpectraMagic™ NX Quality Control Data Software Pack.
- Transmittance Sample Holder Fixture.
- Glass and Plastic Liquid Specimen Cells available in optical path lengths of 2 mm, 10 mm, and 20 mm.
- Opaque Transmittance Calibration Plate.
- Traceable Color Calibration Assessment Tiles.
- Extended 5-meter USB Connection Cable.












There are no reviews yet.